9RSU | pdb_00009rsu

NetF pore structure


ELECTRON MICROSCOPY

Starting Model(s)

Initial Refinement Model(s)
TypeSourceAccession CodeDetails
otherOther 

Refinement

RMS Deviations
KeyRefinement Restraint Deviation
f_dihedral_angle_d4.217
f_angle_d0.427
f_chiral_restr0.041
f_bond_d0.004
f_plane_restr0.003
Sample
NetF
Specimen Preparation
Sample Aggregation StatePARTICLE
Vitrification InstrumentFEI VITROBOT MARK IV
Cryogen NameETHANE
Sample Vitrification Details
3D Reconstruction
Reconstruction MethodSINGLE PARTICLE
Number of Particles278866
Reported Resolution (Å)2.1
Resolution MethodFSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry TypePOINT
Point SymmetryC8
Map-Model Fitting and Refinement
Id1
Refinement Space
Refinement ProtocolOTHER
Refinement Target
Overall B Value
Fitting Procedure
DetailsModelangelo was used to generate the initial model
Data Acquisition
Detector TypeFEI FALCON IV (4k x 4k)
Electron Dose (electrons/Å**2)40
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelTFS KRIOS
Minimum Defocus (nm)1200
Maximum Defocus (nm)2500
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2.7
Imaging ModeBRIGHT FIELD
Specimen Holder ModelFEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification
Calibrated Magnification
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)300
Imaging Details
EM Software
TaskSoftware PackageVersion
PARTICLE SELECTIONcryoSPARC
IMAGE ACQUISITIONEPU
MODEL FITTINGPHENIX
INITIAL EULER ASSIGNMENTcryoSPARC
FINAL EULER ASSIGNMENTcryoSPARC
RECONSTRUCTIONcryoSPARC
MODEL REFINEMENTPHENIX1.21.2_5419
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION