6R7M

Tobacco Mosaic Virus (TMV)


ELECTRON MICROSCOPY
Sample
Tobacco mosaic virus
Specimen Preparation
Sample Aggregation StateHELICAL ARRAY
Vitrification Instrument
Cryogen NameETHANE
Sample Vitrification DetailsCryoWriter
3D Reconstruction
Reconstruction MethodHELICAL
Number of Particles52776
Reported Resolution (Å)1.92
Resolution MethodFSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry TypeHELICAL
Axial SymmetryC1
Axial Rise1.41
Angular Rotation22.03
Map-Model Fitting and Refinement
Id1 (6I5A)
Refinement Space
Refinement Protocol
Refinement Target
Overall B Value
Fitting Procedure
Details
Data Acquisition
Detector TypeGATAN K2 QUANTUM (4k x 4k)
Electron Dose (electrons/Å**2)72
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelFEI TITAN KRIOS
Minimum Defocus (nm)
Maximum Defocus (nm)
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2.7
Imaging ModeBRIGHT FIELD
Specimen Holder ModelFEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification
Calibrated Magnification
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)300
Imaging Details
EM Software
TaskSoftware PackageVersion
PARTICLE SELECTIONRELION3
IMAGE ACQUISITIONSerialEM3.6
MODEL FITTINGUCSF Chimera1.13.1
CLASSIFICATIONRELION3
RECONSTRUCTIONRELION3
MODEL REFINEMENTPHENIX1.14
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION2676