ELECTRON MICROSCOPY
| Sample |
|---|
| T9 icosahedral capsid of a Ty3 retrotransposon |
| Specimen Preparation | |
|---|---|
| Sample Aggregation State | PARTICLE |
| Vitrification Instrument | HOMEMADE PLUNGER |
| Cryogen Name | ETHANE |
| Sample Vitrification Details | The sample was applied onto glow-discharged C-flat (Protochips Inc.) holey carbon grids. The grids were blotted from the back side for 11 seconds at r ... |
| 3D Reconstruction | |
|---|---|
| Reconstruction Method | SINGLE PARTICLE |
| Number of Particles | 1236 |
| Reported Resolution (Å) | 7.5 |
| Resolution Method | FSC 0.143 CUT-OFF |
| Other Details | |
| Refinement Type | |
| Symmetry Type | POINT |
| Point Symmetry | I |
| Map-Model Fitting and Refinement | |||||
|---|---|---|---|---|---|
| Id | 1 | ||||
| Refinement Space | REAL | ||||
| Refinement Protocol | RIGID BODY FIT | ||||
| Refinement Target | Cross-correlation coefficient | ||||
| Overall B Value | |||||
| Fitting Procedure | |||||
| Details | |||||
| Data Acquisition | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Detector Type | FEI FALCON II (4k x 4k) | ||||||||
| Electron Dose (electrons/Å**2) | 20 | ||||||||
| Imaging Experiment | 1 |
|---|---|
| Date of Experiment | |
| Temperature (Kelvin) | |
| Microscope Model | FEI TITAN KRIOS |
| Minimum Defocus (nm) | 1000 |
| Maximum Defocus (nm) | 3500 |
| Minimum Tilt Angle (degrees) | |
| Maximum Tilt Angle (degrees) | |
| Nominal CS | 2.7 |
| Imaging Mode | BRIGHT FIELD |
| Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
| Nominal Magnification | 75000 |
| Calibrated Magnification | |
| Source | FIELD EMISSION GUN |
| Acceleration Voltage (kV) | 300 |
| Imaging Details | Cryo-grids containing purified PR- Ty3 particles were imaged in a Titan Krios electron microscope equipped with a Falcon II direct electron detector, operated at 300 kV. Images were collected with a nominal magnification of 75000, giving a pixel size of 1.08 A. Images were collected in integrating mode with a total electron dose of 20 e/A2. The range of applied defocus values was between -1.0 um and -3.5 um. |
| EM Software | ||
|---|---|---|
| Task | Software Package | Version |
| PARTICLE SELECTION | EMAN2 | |
| IMAGE ACQUISITION | EPU | |
| CTF CORRECTION | CTFFIND | 4 |
| MODEL FITTING | UCSF Chimera | 1.13 |
| INITIAL EULER ASSIGNMENT | RELION | |
| FINAL EULER ASSIGNMENT | RELION | |
| CLASSIFICATION | RELION | |
| RECONSTRUCTION | RELION | |
| Image Processing | ||||
|---|---|---|---|---|
| CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
| PHASE FLIPPING AND AMPLITUDE CORRECTION | 1727 | |||