6Z0U

CryoEM structure of the Chikungunya virus nsP1 complex


ELECTRON MICROSCOPY

Refinement

RMS Deviations
KeyRefinement Restraint Deviation
f_dihedral_angle_d7.6905
f_angle_d0.6938
f_chiral_restr0.0504
f_bond_d0.0069
f_plane_restr0.0054
Sample
Chikungunya NsP1 capping pore complex
Specimen Preparation
Sample Aggregation StatePARTICLE
Vitrification InstrumentLEICA EM CPC
Cryogen NameETHANE-PROPANE
Sample Vitrification Details3ul of sample was spotted onto the grid and hand blotted prior to plunge freezing.
3D Reconstruction
Reconstruction MethodSINGLE PARTICLE
Number of Particles47338
Reported Resolution (Å)2.9
Resolution MethodFSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry TypePOINT
Point SymmetryD12
Map-Model Fitting and Refinement
Id1
Refinement SpaceREAL
Refinement ProtocolAB INITIO MODEL
Refinement TargetCC
Overall B Value
Fitting Procedure
Details
Data Acquisition
Detector TypeGATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å**2)38.5
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelFEI TITAN KRIOS
Minimum Defocus (nm)1000
Maximum Defocus (nm)2500
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2.7
Imaging ModeBRIGHT FIELD
Specimen Holder ModelFEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification165000
Calibrated Magnification
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)300
Imaging Details
EM Software
TaskSoftware PackageVersion
IMAGE ACQUISITIONEPU1.5.0.32
CTF CORRECTIONCTFFIND4.1
MODEL FITTINGCoot0.8.9.2
MODEL REFINEMENTPHENIX1.17.1
INITIAL EULER ASSIGNMENTRELION3.0.8
FINAL EULER ASSIGNMENTRELION3.0.8
CLASSIFICATIONRELION3.0.8
RECONSTRUCTIONRELION3.0.8
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTIONCTF was estimated from movie frames aligned without dose-weighting. Per particle CTF correction was also performed with polished particles.263415Autopicking following generation of templates from 2D class averages