Compact, ligand-free state of Manduca sexta soluble guanylate cyclase mutant beta C122S
ELECTRON MICROSCOPY
Refinement
| RMS Deviations | |
|---|---|
| Key | Refinement Restraint Deviation |
| f_dihedral_angle_d | 8.469 |
| f_angle_d | 0.49 |
| f_chiral_restr | 0.039 |
| f_plane_restr | 0.006 |
| f_bond_d | 0.002 |
| Sample |
|---|
| Ligand-free Manduca sexta soluble guanylase cyclase variant |
| Specimen Preparation | |
|---|---|
| Sample Aggregation State | PARTICLE |
| Vitrification Instrument | FEI VITROBOT MARK IV |
| Cryogen Name | ETHANE |
| Sample Vitrification Details | Cryo-EM samples were prepared by applying 3 ul to a glow-discharged Quantifoil R1.2/1.3 holey-carbon cryo-EM grid. The grid was blotted for 4 s with W ... |
| 3D Reconstruction | |
|---|---|
| Reconstruction Method | SINGLE PARTICLE |
| Number of Particles | 264227 |
| Reported Resolution (Å) | 3.2 |
| Resolution Method | FSC 0.143 CUT-OFF |
| Other Details | During refinement it became clear that there was minor interdomain flexibility within the 264,227 particles. In particular, the catalytic domain was a ... |
| Refinement Type | |
| Symmetry Type | POINT |
| Map-Model Fitting and Refinement | |||||
|---|---|---|---|---|---|
| Id | 1 | ||||
| Refinement Space | REAL | ||||
| Refinement Protocol | FLEXIBLE FIT | ||||
| Refinement Target | |||||
| Overall B Value | |||||
| Fitting Procedure | |||||
| Details | Refinement was performed using iterative rounds of Phenix real space refinement and manual modeling in Coot. Phenix refinement was performed for separ ... | ||||
| Data Acquisition | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Detector Type | FEI FALCON I (4k x 4k) | ||||||||
| Electron Dose (electrons/Å**2) | 1.25 | ||||||||
| Imaging Experiment | 1 |
|---|---|
| Date of Experiment | |
| Temperature (Kelvin) | |
| Microscope Model | TFS KRIOS |
| Minimum Defocus (nm) | 500 |
| Maximum Defocus (nm) | 1500 |
| Minimum Tilt Angle (degrees) | |
| Maximum Tilt Angle (degrees) | |
| Nominal CS | 2.7 |
| Imaging Mode | BRIGHT FIELD |
| Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
| Nominal Magnification | |
| Calibrated Magnification | |
| Source | FIELD EMISSION GUN |
| Acceleration Voltage (kV) | 300 |
| Imaging Details |
| EM Software | ||
|---|---|---|
| Task | Software Package | Version |
| PARTICLE SELECTION | cryoSPARC | |
| IMAGE ACQUISITION | EPU | |
| CTF CORRECTION | cryoSPARC | |
| MODEL FITTING | UCSF ChimeraX | |
| INITIAL EULER ASSIGNMENT | cryoSPARC | |
| FINAL EULER ASSIGNMENT | cryoSPARC | |
| CLASSIFICATION | cryoSPARC | |
| RECONSTRUCTION | cryoSPARC | |
| MODEL REFINEMENT | PHENIX | 1.21.1_5286 |
| MODEL REFINEMENT | Coot | |
| Image Processing | ||||
|---|---|---|---|---|
| CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
| PHASE FLIPPING AND AMPLITUDE CORRECTION | 1226641 | Autopicking was used to produce a stack of 1,226,641 particles. Particles showing poor alignment or broken complexes were removed using a series of 2D classification steps, leaving a particle stack of 368,563 particles that resembled sGC dimers. | ||