6S6B

Type III-B Cmr-beta Cryo-EM structure of the Apo state


ELECTRON MICROSCOPY

Refinement

RMS Deviations
KeyRefinement Restraint Deviation
f_dihedral_angle_d7.456
f_angle_d0.804
f_chiral_restr0.053
f_bond_d0.008
f_plane_restr0.005
Sample
Type III-B Cmr-beta binary complex, crRNA-bound
Specimen Preparation
Sample Aggregation StatePARTICLE
Vitrification InstrumentFEI VITROBOT MARK IV
Cryogen NameETHANE
Sample Vitrification Details3-4 s blotting before plunging
3D Reconstruction
Reconstruction MethodSINGLE PARTICLE
Number of Particles69043
Reported Resolution (Å)2.75
Resolution MethodFSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry TypePOINT
Point SymmetryC1
Map-Model Fitting and Refinement
Id1
Refinement SpaceREAL
Refinement ProtocolAB INITIO MODEL
Refinement TargetCross-correlation coefficient
Overall B Value
Fitting Procedure
Details
Data Acquisition
Detector TypeFEI FALCON III (4k x 4k)
Electron Dose (electrons/Å**2)40
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelFEI TITAN KRIOS
Minimum Defocus (nm)1700
Maximum Defocus (nm)2600
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2.1
Imaging ModeBRIGHT FIELD
Specimen Holder ModelFEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification96000
Calibrated Magnification
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)300
Imaging Details
EM Software
TaskSoftware PackageVersion
IMAGE ACQUISITIONEPU
CTF CORRECTIONGctf1.06
MODEL FITTINGUCSF Chimera
MODEL FITTINGCoot
MODEL FITTINGPHENIX
INITIAL EULER ASSIGNMENTRELION3.0
FINAL EULER ASSIGNMENTcisTEM1.0
CLASSIFICATIONcisTEM1.0
RECONSTRUCTIONcisTEM1.0
MODEL REFINEMENTPHENIX
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION614000