5KG8

Rigor myosin X co-complexed with an actin filament


ELECTRON MICROSCOPY
Sample
Actin filament decorated by the myosin X motor domain
Specimen Preparation
Sample Aggregation StateFILAMENT
Vitrification InstrumentHOMEMADE PLUNGER
Cryogen NameETHANE
Sample Vitrification Details
3D Reconstruction
Reconstruction MethodHELICAL
Number of Particles57927
Reported Resolution (Å)9.1
Resolution MethodFSC 0.143 CUT-OFF
Other DetailsSymmetry was not applied in this reconstruction. The provided symmetry parameters are nominal, and were not actually used.
Refinement Type
Symmetry TypeHELICAL
Axial SymmetryC1
Axial Rise27.44
Angular Rotation167.1
Data Acquisition
Detector TypeGATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å**2)50
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelFEI TECNAI F20
Minimum Defocus (nm)1000
Maximum Defocus (nm)5000
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2
Imaging ModeBRIGHT FIELD
Specimen Holder ModelGATAN 626 SINGLE TILT LIQUID NITROGEN CRYO TRANSFER HOLDER
Nominal Magnification
Calibrated Magnification26780
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)200
Imaging Details
EM Software
TaskSoftware PackageVersion
PARTICLE SELECTIONboxer
IMAGE ACQUISITIONSerialEM
CTF CORRECTIONCTFFIND3
INITIAL EULER ASSIGNMENTSPARX
FINAL EULER ASSIGNMENTFREALIGN8.06
RECONSTRUCTIONFREALIGN8.06
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION