3J2X

Electron Cryo-microscopy of Chikungunya VLP in complex with neutralizing antibody Fab m242


ELECTRON MICROSCOPY
Sample
Electron Cryo-microscopy of Chikungunya VLP in complex with neutralizing antibody Fab m242
Sample Components
m242 antibody Fab fragment
Chikungunya virus like particle
Specimen Preparation
Sample Aggregation StatePARTICLE
Vitrification InstrumentGATAN CRYOPLUNGE 3
Cryogen NameETHANE
Sample Vitrification DetailsBlot for 2 seconds before plunging into liquid ethane (GATAN CRYOPLUNGE 3)
3D Reconstruction
Reconstruction MethodSINGLE PARTICLE
Number of Particles1728
Reported Resolution (Å)15.6
Resolution MethodFSC 0.5 CUT-OFF
Other Details(Single particle details: The particles were selected using e2boxer) (Single particle--Applied symmetry: I)
Refinement Type
Symmetry TypePOINT
Point SymmetryI
Map-Model Fitting and Refinement
Id1 (4GQ8)
Refinement SpaceREAL
Refinement ProtocolRIGID BODY FIT
Refinement TargetSumf
Overall B Value
Fitting Procedure
DetailsREFINEMENT PROTOCOL--rigid body
Data Acquisition
Detector Type
Electron Dose (electrons/Å**2)20
Imaging Experiment1
Date of Experiment2011-01-01
Temperature (Kelvin)
Microscope ModelFEI/PHILIPS CM300FEG/HE
Minimum Defocus (nm)
Maximum Defocus (nm)
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2
Imaging ModeBRIGHT FIELD
Specimen Holder ModelGATAN LIQUID NITROGEN
Nominal Magnification38000
Calibrated Magnification39190
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)300
Imaging Details
EM Software
TaskSoftware PackageVersion
MODEL FITTINGEMfit
RECONSTRUCTIONEMAN2
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
Each Micrograph